Share:
Share this content in WeChat
X
Clinical Article
Evaluate measurement value of liver volume before pediatric living donor liver transplantation by MRI Argus
PENG Pai  MU Xue-tao  YOU Bo  ZHANG Yu  Yin Yuanyuan  DONG Yu-ru  LI Wei  WANG Hong 

DOI:10.12015/issn.1674-8034.2016.08.008.


[Abstract] Objective: To evaluate the accuracy of the application of MRI Argus assessment in liver volume before pediatric living donor liver transplantation (PLDLT).Materials and Methods: Eleven PLDLT donor livers underwent enhanced MRI scanning. Donor liver volumes were calculated with importing venous phase reconstruction data to Argus software, correlation analysis was used with actual removal of the liver weight during the operation.Results: Paired t-test showed there was statistically significant difference` (P<0.01) between liver volume measured by MR Argus software (357.00±72.37) ml and real liver weight in intraoperative actual removal (284.45±61.27) g. Person correlation test showed liver volume measured by MR Argus software had a significantly positive correlation with real liver weight (r=0.982, P<0.001). Graft weight could be expressed in the equation: W real liver weight (g) = 0.832 × V liver volume measured by MR Argus-12.46.Conclusion: MRI Argus software is helpful to estimate graft volume before surgery, utilizing the regression equation can accurately forecast the quality of liver transplantation, and providing assessment for preoperative graft volume.
[Keywords] Magnetic resonance imaging;Child;Liver transplantation;Volumetric measurement

PENG Pai Department of MRI, the General Hospital of Chinese People’s Armed Police Forces, Beijing 100039, China; Xuzhou Medical College, Xuzhou 221000, China

MU Xue-tao Department of MRI, the General Hospital of Chinese People’s Armed Police Forces, Beijing 100039, China

YOU Bo Institute of Organ Transplantation, the General Hospital of Chinese Peopleˊs Armed Police Forces, Beijing 100039, China

ZHANG Yu Department of MRI, the General Hospital of Chinese People’s Armed Police Forces, Beijing 100039, China

Yin Yuanyuan Department of MRI, the General Hospital of Chinese People’s Armed Police Forces, Beijing 100039, China

DONG Yu-ru Department of MRI, the General Hospital of Chinese People’s Armed Police Forces, Beijing 100039, China

LI Wei Institute of Organ Transplantation, the General Hospital of Chinese Peopleˊs Armed Police Forces, Beijing 100039, China

WANG Hong* Department of MRI, the General Hospital of Chinese People’s Armed Police Forces, Beijing 100039, China

*Correspondence to: Wang H, E-mail: wanghongmri@sina.com

Conflicts of interest   None.

ACKNOWLEDGMENTS  This work was part of High technology National development and reform commission 2013 No. 2140
Received  2016-01-16
Accepted  2016-06-30
DOI: 10.12015/issn.1674-8034.2016.08.008
DOI:10.12015/issn.1674-8034.2016.08.008.

[1]
Organ Procurement and Transplantation Network. OPTN/SRTR 2012 Annual Data Report: liver. base on OPTN data as of 2014. [2016-01-10]. http://onlinelibrary.wiley.com/doi/10.1111/ajt.12581/pdf.
[2]
European Liver Transplant Registry. Survival according to children recipient age, based on ELTR data as of December, 2010. (2011-12-25)[2016-01-10]. http://www.eltr.org/Pediatric-transplantation.html.
[3]
中国肝移植注册(CLTR). 2011中国肝移植年度科学报告. 2013. [2016-01-10]. http://www.cltr.org/pages/datainfo/datainfo_apparatus.jsp?subType=11.
[4]
潘晶晶,张静,王海,等.一站式MRI在活体肝移植术前评估中的价值.中国临床医学影像杂志, 2014, 25(3): 627-629.
[5]
赖彦华,董家鸿,段伟东,等.精准肝脏外科技术在活体肝移植供肝切取中的应用.中华外科杂志, 2015, 53(5): 328-334.
[6]
Kokudo T, Hasegawa K, Uldry E, et al. A new formula for calculating standard liver volume for living donor liver transplantation without using body weight. Hepatol, 2015, 63(4): 848-854.
[7]
Herden U, Wischhusen F, Heinemann A, et al. A formula to calculate the standard liver volume in children and its application in pediatric liver transplantation. Transpl Int, 2013, 26(12): 1217-1224.
[8]
Li YC, Hu Y, Zhang MM, et al. Usage of 64-detector-row spiral computed tomography volumetry in preoperative volume prediction in living donor liver transplantation in children. Pediatr Surg Int, 2011, 27(5): 445-449.
[9]
Bertell R, Ehrle LH, Schmitz-Feuerhake I. Pediatric CT research elevates public health concerns: low-dose radiation issues are highly politicized. Health Serv, 2007, 37(3): 419-439.
[10]
李晓会,靳囡,白芝兰,等.非离子型碘造影剂过敏试验的应用体会.实用放射学杂志, 2010, 26(2): 294-296.
[11]
张水兴,刘于宝,梁长虹,等.肝脏磁共振对比剂的过去、现在和未来.磁共振成像, 2012, 3(1): 56-60.
[12]
穆学涛,王宏,吴春楠,等. Argus和手工测量法对活体肝移植供体术后肝体积测量的对比研究.医疗卫生装备, 2012, 33(1): 70-72.
[13]
Satou S, Sugawara Y, Tamura S, et al. Discrepancy between estimated and actual weight of partial liver graft from living donors. Hepatobiliary Pancreat Sci, 2011, 18(4): 586-591.
[14]
Yoneyama T, Asonuma K, Okajima H, et al. Coefficient factor for graft weight estimation from preoperative computed tomography volumetry in living donor liver transplantation. Liver Transpl, 2011, 17(4): 369-372.

PREV The diagnostic value of CT and MRI in spinal tuberculosis
NEXT The effect of intentionality on reactive aggression: a rs-fMRI study on individual differences
  



Tel & Fax: +8610-67113815    E-mail: editor@cjmri.cn